EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectrum Measurements
R&S
®
ESR
431User Manual 1175.7068.02 ─ 12
In the following examples, a signal generator is used as a signal source. The RF output
of the signal generator is connected to the RF input of R&S ESR.
If a 64 MHz signal is required for the test setup, as an alternative to the signal genera-
tor, the internal 64 MHz reference generator can be used:
1. Switch on the internal reference generator.
a) Press the SETUP key.
b) Press the "More" key.
c) Press the "Service" softkey.
d) Press the "Input RF/Cal" softkey, until "Cal" is highlighted.
The internal 64 MHz reference generator is now on. The R&S ESR's RF input
is switched off.
2. Switch on the RF input again for normal operation of the R&S ESR. Two ways are
possible:
a) Press the PRESET key.
or:
b) Press the SETUP key.
c) Press the "Service" softkey.
d) Press the "Input RF/Cal" softkey, until "RF" is highlighted.
The internal signal path of the R&S ESR is switched back to the RF input in order
to resume normal operation.
6.4.2 Measurement of Harmonics
Signal generator settings (e.g. R&S SMW):
Frequency: 128 MHz
Level: - 25 dBm
Procedure on the R&S ESR:
1. Set the R&S ESR to its default state by pressing the [PRESET] key.
2. Set the center frequency to 128 MHz and the span to 100 kHz.
3. Switch on the marker by pressing the [MKR] key.
The marker is positioned on the trace maximum.
4. Set the measured signal frequency and the measured level as reference values.
a) Press the [MKR FUNC] key
b) Press the "Ref Fixed" softkey.
The position of the marker becomes the reference point. The reference point
level is indicated by a horizontal line, the reference point frequency with a verti-
cal line. At the same time, the delta marker 2 is switched on.
Advanced Measurement Examples

Table of Contents

Related product manuals