Remote Control
R&S
®
ESR
792User Manual 1175.7068.02 ─ 12
Configuring Statistical Measurements...................................................................792
Retrieving Results.................................................................................................793
Using Gate Ranges for Statistical Measurements................................................ 794
Scaling the Diagram..............................................................................................796
Configuring Statistical Measurements
The following commands configure the measurement.
CALCulate<n>:STATistics:APD[:STATe]............................................................................792
CALCulate<n>:STATistics:CCDF[:STATe]......................................................................... 792
CALCulate<n>:STATistics:NSAMples...............................................................................793
CALCulate<n>:STATistics:APD[:STATe] <State>
This command switches on or off the measurement of amplitude distribution (APD). On
activating this function, the CCDF measurement is switched off.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:STAT:APD ON
Switches on the APD measurement.
Manual operation: See "APD" on page 224
See "APD" on page 316
CALCulate<n>:STATistics:CCDF[:STATe] <State>
This command switches on or off the measurement of the complementary cumulative
distribution function (CCDF). On activating this function, the APD measurement is
switched off.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:STAT:CCDF ON
Switches on the CCDF measurement.
Manual operation: See "CCDF" on page 224
See "CCDF" on page 321