Receiver Mode
R&S
®
ESR
220User Manual 1175.7068.02 ─ 12
5.4.3.9 Activating/Deactivating a Limit Line
Prerequisites:
The x- and y-units of limit line and current measurement setting have to be compatible.
For details refer to Chapter 5.4.3.3, "Limit Lines (Frequency/Time Lines)", on page 214.
The limit line has to consist of 2 or more data points.
1. In the "Select Limit Line" dialog box, select the limit line you want to activate/deacti-
vate. For details see also Chapter 5.4.3.5, "Selecting a Limit Line", on page 215.
2. To activate or deactivate a limit line for a trace, press the "Select Traces to check"
on page 210 softkey and select or deselect the trace(s) to which this limit line
applies.
3. To deactivate the limit line for all traces, press the "Deselect All" on page 210 soft-
key.
Common Analysis Functions