Remote Control
R&S
®
ESR
953User Manual 1175.7068.02 ─ 12
Manual operation: See "Defining characteristics of a transducer factor"
on page 508
[SENSe:]CORRection:TRANsducer:SELect <FileName>
This command creates or selects a transducer factor.
Parameters:
<FileName> String containing the name of the transducer factor.
If the name does not exist yet, the R&S ESR creates a trans-
ducer factor by that name.
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Using transducer factors or transducer sets" on page 508
See "Defining characteristics of a transducer factor"
on page 508
[SENSe:]CORRection:TRANsducer[:STATe]
<State>
This command turns a transducer factor on and off.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
You can turn on up to 8 transducer factors at the same time.
Parameters:
<State> ON | OFF
*RST: OFF
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Using transducer factors or transducer sets" on page 508
[SENSe:]CORRection:TRANsducer:UNIT <Unit>
This command selects the unit of a transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
Parameters:
<Unit> String containing one of the following units:
DB | DBM | DBMV | DBUV | DBUV/M | DBUA DBUA/M | DBPW |
DBPT
*RST: DB
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Defining characteristics of a transducer factor"
on page 508
Using Transducers