EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
ESR
953User Manual 1175.7068.02 ─ 12
Manual operation: See "Defining characteristics of a transducer factor"
on page 508
[SENSe:]CORRection:TRANsducer:SELect <FileName>
This command creates or selects a transducer factor.
Parameters:
<FileName> String containing the name of the transducer factor.
If the name does not exist yet, the R&S ESR creates a trans-
ducer factor by that name.
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Using transducer factors or transducer sets" on page 508
See "Defining characteristics of a transducer factor"
on page 508
[SENSe:]CORRection:TRANsducer[:STATe]
<State>
This command turns a transducer factor on and off.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
You can turn on up to 8 transducer factors at the same time.
Parameters:
<State> ON | OFF
*RST: OFF
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Using transducer factors or transducer sets" on page 508
[SENSe:]CORRection:TRANsducer:UNIT <Unit>
This command selects the unit of a transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
Parameters:
<Unit> String containing one of the following units:
DB | DBM | DBMV | DBUV | DBUV/M | DBUA DBUA/M | DBPW |
DBPT
*RST: DB
Example: See "Designing a transducer factor" on page 949.
Manual operation: See "Defining characteristics of a transducer factor"
on page 508
Using Transducers

Table of Contents