EasyManuals Logo

R&S ESR Series User Manual

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #957 background imageLoading...
Page #957 background image
Remote Control
R&S
®
ESR
954User Manual 1175.7068.02 ─ 12
[SENSe:]CORRection:TRANsducer:VIEW <State>
This command includes or excludes a figure of a transducer factor in the diagram.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CORR:TRAN:SEL 'FACTOR'
CORR:TRAN:VIEW ON
Selects and displays a transducer factor.
[SENSe:]CORRection:TSET:BREak
<State>
This command turns a transducer break on and off.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Parameters:
<State> ON | OFF
*RST: OFF
Example: See "Managing a transducer set" on page 950.
Manual operation: See "Managing transducer sets" on page 510
[SENSe:]CORRection:TSET:CATalog?
This command queries all transducer sets stored on the R&S ESR.
Return values:
<UsedDiskSpace> Size of all available files containing transducer factors in byte.
<FreeDiskSpace> Remaining disk space in bytes.
<FileInfo> String containing the file information.
'<NameFileN>,<SizeFileN>'
<NameFileN>
Name of the file.
<SizeFileN>
Size of the file in bytes.
Example:
CORR:TSET:CAT?
would return, e.g.
3298,34482896896,'SET1,1520','SET2,1756'
Usage: Query only
Using Transducers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals