Remote Control
R&S
®
ESR
955User Manual 1175.7068.02 ─ 12
[SENSe:]CORRection:TSET:COMMent <Comment>
This command defines a comment for a transducer set.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Parameters:
<Comment> String containing the comment.
Example: See "Managing a transducer set" on page 950.
Manual operation: See "Managing transducer sets" on page 510
[SENSe:]CORRection:TSET:DELete
This command deletes a transducer set.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Example:
CORR:TSET:SEL 'Transducer Set'
CORR:TSET:DEL
Deletes the transducer set.
Usage: Event
Manual operation: See "Managing transducer sets" on page 510
[SENSe:]CORRection:TSET:RANGe<range> <Frequency>, <Frequency>,
<FileName>
This command selects a set of transducer factors used for a particular frequency
range.
Ranges 1 to 10 must be sent in ascending order.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Suffix:
<range>
.
1...10
Selects the range.
Parameters:
<Frequency> Defines the start frequency of the frequency range.
<Frequency> Defines the stop frequency of the frequency range.
<FileName> String containing the name of the transducer factor.
Note that you can assign up to eight transducer factors to a par-
ticular frequency range. In that case, add additional strings con-
taining the file name after the first one separated by comma.
Example: See "Managing a transducer set" on page 950.
Manual operation: See "Managing transducer sets" on page 510
Using Transducers