Remote Control
R&S
®
ESR
780User Manual 1175.7068.02 ─ 12
POS
maximum peak detector
QPE
quasipeak detector
*RST: OFF
Manual operation: See "Marker Configuration" on page 303
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector <Detector>
This command selects the detector for a specific marker during the final measurement.
If the marker hasn't been active yet, the command also turns the marker on.
Suffix:
<n>
.
irrelevant
<m> marker
Parameters:
<Detector> OFF
turns the marker off
AVER
average detector
CAV
CISPR Average detector
CRMS
RMS Average detector
POS
maximum peak detector
QPE
quasipeak detector
*RST:
OFF
Manual operation: See "Marker Configuration" on page 303
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl <Time>
This command defines the dwell time during the final measurement.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<Time> Range: min value to max value
*RST: 1 s
Default unit: seconds
Manual operation: See "Marker Configuration" on page 303
See "Dwell Time" on page 305
Remote Commands in Spectrum Analyzer Mode