Remote Control
R&S
®
ESR
782User Manual 1175.7068.02 ─ 12
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?
This command queries the vertical distance from the deltamarker position to the limit
line. The unit is dB.
If the deltamarker has been assigned to a different trace than the limit line, the com-
mand returns -200.
Suffix:
<n>
.
irrelevant
<m> 1...16
deltamarker
<k> 1...8
limit line
Return values:
<Amplitude> Vertical distance to the limit line in dB.
Example:
CALC:DELT3:FUNC:FME:LIM2:DELT?
Queries the distance of deltamarker 3 to the second limit line.
Usage: Query only
Manual operation: See "Marker Table" on page 305
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:RESult? <Result>
This command queries the result of the EMI measurement at the deltamarker position.
Suffix:
<n>
.
irrelevant
<m> 1...16
marker
Return values:
<Result> Power level in dB related to the reference marker you have set
for that delta marker.
Example:
CALC:DELT3:FUNC:FME:RES?
Queries the result of delta marker 3.
Usage: Query only
Manual operation: See "Marker Table" on page 305
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?
<Condition>
This command queries the condition of a marker position in relation to a certain limit
line.
Suffix:
<n>
.
irrelevant
Remote Commands in Spectrum Analyzer Mode