Remote Control
R&S
®
ESR
789User Manual 1175.7068.02 ─ 12
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Example:
CALC:MARK:FUNC:HARM:PRES
Optimizes the device settings for the harmonic measurement.
Usage: Event
Manual operation: See "Adjust Settings" on page 341
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe] <State>
This command turns the harmonic distortion measurement on and off.
Note the following:
If you perform the measurement in the frequency domain, the search range for the
frequency of the first harmonic, whose power is determined, is defined by the last
span.
If you perform the measurement in the time domain, the current center frequency is
used as the frequency of the first harmonic. Thus, the frequency search is
bypassed. The first harmonic frequency is set by a specific center frequency in
zero span before the harmonic measurement is started.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
Manual operation: See "Harmonic Distortion" on page 225
See "Harmonic Distortion" on page 340
UNIT:THD <Mode>
This command selects the unit for THD measurements.
Parameters:
<Mode> DB | PCT
*RST: DB
Example:
UNIT:THD PCT
Remote Commands in Spectrum Analyzer Mode