Remote Control
R&S
®
ESR
813User Manual 1175.7068.02 ─ 12
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Return values:
<PulsePower> Lowest power level of the pulse power measurement.
The unit is dBm.
Example:
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of
an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,
5us,MEAN,20;*OPC
Configuring and starting the measurement
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult:MIN?
Usage:
Query only
[SENSe:]MPOWer[:SEQuence] <Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <#OfPulses>
This command configures and initiates the pulse power measurement.
The R&S ESR caches all measurement parameters that you can set with this com-
mand. If you use the command repeatedly, the R&S ESR only changes those settings
that you have actually changed before initiating the measurement. Thus, measurement
times are kept as low as possible.
If you synchronize the measurement with *OPC, the R&S ESR produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see .
To reduce the setting time, the setup is performed simultaneously for all selected
parameters.
The query adapts the instrument settings, performs the defined number of measure-
ments and outputs the measurement results list.
This measurement is not compatible with other measurements, especially as far as
marker functions, adjacent-channel measurement or statistics are concerned. The cor-
responding functions are therefore automatically switched off. In return, incompatible
commands automatically deactivate the multi burst power function.
Remote Commands in Spectrum Analyzer Mode