Remote Control
R&S
®
ESR
911User Manual 1175.7068.02 ─ 12
Manual operation: See "Generator Frequency (Numerator/Denominator/Offset)"
on page 489
SOURce<n>:EXTernal<generator>:FREQuency:SWEep[:STATe] <State>
This command activates or deactivates the frequency sweep for the selected genera-
tor.
This command is requires the optional External Tracking Generator.
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
external generator
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SOUR:EXT1:FREQ:SWE ON
Activates the frequency sweep for external generator 1.
SOURce<n>:EXTernal<generator>:POWer[:LEVel] <Level>
This command sets the output power of the selected generator.
This command is only valid for the External Tracking Generator option (R&S FSV-B10).
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Level> <numeric value>
*RST: -20 dBm
Example:
SOUR:EXT:POW -30dBm
Sets the generator level to -30 dBm
Mode: A, NF
Manual operation: See "Source Power" on page 489
SOURce<n>:EXTernal<generator>:ROSCillator[:SOURce] <Source>
This command controls selection of the reference oscillator for the external tracking
generator (requires option R&S FSV-B10).
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Remote Commands to Control the Tracking Generator