Remote Control
R&S
®
ESR
928User Manual 1175.7068.02 ─ 12
*RST: ALL
Example:
INST:COUP:ATT ALL
Synchronizes the attenuation.
INSTrument:COUPle:BANDwidth <State>
This command turns synchronization of the resolution bandwidth (and filter type)
between measurement channels on and off.
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns on synchronization.
*RST: NONE
Example:
INST:COUP:BWID ALL
Synchronizes the resolution bandwidth.
INSTrument:COUPle:CENTer <State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: ALL
Example:
INST:COUP:CENT ALL
Synchronizes the receiver (or center) frequency.
INSTrument:COUPle:DEMod <State>
This command turns synchronization of the audio demodulator configuration between
measurement channels on and off.
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:DEM ALL
Synchronizes the audio demodulator configuration.
System Configuration