6-12
IM DLM6054-01EN
Setting Patterns to Use as Conditions (Pattern)
Press the Pattern soft key to display the following menu.
Use the jog shuttle or the SET knob to
move from bit to bit, and use the SET knob
or the soft keys to set the bit pattern.
Explanation
You can use this feature to make the DL6000/DLM6000 trigger on the trigger source edge when the
conditions of the signals other than the trigger source meet the specified qualifications.
The settings for this feature vary depending on whether the trigger source is an analog or logic
waveform.
Triggering on an External Signal or a Channel from CH1 to CH4
(Edge Qualified)
Trigger Source
You can set the trigger source to a channel from CH1 to CH4, or EXT.
Triggering on an External Signal
To use the external signal that the DL6000/DLM6000 receives through the rear-panel TRIG IN
terminal as the trigger source, set SOURCE to EXT.
Trigger Slope
You can select how the DL6000/DLM6000 will trigger on the trigger source signal.
The DL6000/DLM6000 triggers when the trigger source changes from a level below
the trigger level to a level above the trigger level (rising).
The DL6000/DLM6000 triggers when the trigger source changes from a level above
the trigger level to a level below the trigger level (falling).
Enter The DL6000/DLM6000 triggers when the trigger source level enters the specified
voltage range (when the window comparator is on).
Exit The DL6000/DLM6000 triggers when the trigger source level exits from the specified
voltage range (when the window comparator is on).
Qualification
For each signal, you can set the condition that must be met for the trigger to be valid to H, L, or X
H The signal level must be high.
L The signal level must be low.
IN The signal level must be within the specified voltage range (when the trigger source is
a channel from CH1 to CH4 and the window comparator is on).
OUT The signal level must be outside the specified voltage range (when the trigger source is
a channel from CH1 to CH4 and the window comparator is on).
X No conditions have to be met (“don’t care”).
Logic
You can set the logic for all of the conditions to AND or OR.
AND The DL6000/DLM6000 triggers when all of the specified conditions are met.
OR The DL6000/DLM6000 triggers when any of the specified conditions is met.
6.4 Triggering on an Edge Condition