Triggering
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Index
App
6-49
IM DLM6054-01EN
For Logic Sources
Set the trigger source type
to Logic.
Select the active state.
Select the data source.
Select the clock source.
Turns clock source
synchronization on and off
(the same as the Clock setting
in the Condition tab)
State condition settings
Select rising or falling.
Turns state condition
comparison on and off
Set the state conditions
(when CS is set to ON).
Select the latch source.
(Select a bit from A0 to A7,
B0 to B7, C0 to C7, D0 to D7,
or select X to not specify a latch.)
Select rising or falling.
When Clock Is Set to ON
When Clock Is Set to OFF
Set the bit rate.
Explanation
You can use this feature to trigger on a serial pattern signal. The DL6000/DLM6000 compares the
source signal with the data pattern that you set and triggers when the two match.
Data, Clock, CS, and Latch Sources (Select)
You can set the sources for the data, clock, CS, and latch signals to analog signals or logic signal bits.
When you choose analog signals, you can select the sources from CH1 to CH4.
When you choose the logic signal, you can select the sources from bits A0 to A7, B0 to B7, and D0 to
D7. (On 16-bit models, you can select from bits A0 to A7 and C0 to C7.)
Clock
Clock Setting
You can select whether or not to sample the data source in sync with the selected clock source.
ON The DL6000/DLM6000 samples in sync with the clock source.
OFF The DL6000/DLM6000 does not synchronize to the clock source.
Sample Timing
You can select which edge of the clock source to sample data patterns on.
On the rising edge
On the falling edge
6.13 Triggering on a Serial Pattern Signal