Code Message and Corrective Action Section
663 Retrievable settings does not exist.
The settings is created by either Initialize or Auto Setup.
—
664 Failed to execute statistical measurement.
Waveform data may not exist. In Cycle statistic mode, improper setting may result in failure
to recognize the cycle.
10.3
665 Search target data does not exist.
The search after analysis is completed.
—
666 Improper action setting.
The saved data type is either Waveform group or Analysis group. This can be assigned from
File menu.
Chapter 13
667 Retrievable data not found. —
669 Sending E-Mail.
Wait until it is completed.
—
670 The corresponding field was not found. —
674 Cannnot store because the data is locked.
Release the lock through Store Detail.
4.6
675 Serial bus automatic setting is in progress.
Please wait.
Separate
manual
*
* IM DLM6054-51EN
Setting Errors
Code Message and Corrective Action Section
800 Improper Date/Time setting. 3.7
801 Not allowed unless waveforms are shown.
Display waveforms.
Chapter 5
802 Source waveforms do not exist.
Display source waveforms.
Chapter 7
803 Zone waveforms do not exist. Chapter 7
804 Illegal expression. Chapter 9
805 Not allowed in Mask test mode.
Turn off the Mask test mode.
10.8
806 Invalid bit assignment in the logic group. 5.2
807 Unable to enable the trigger conditions.
Set the clock source to another group or assign bits to the group.
5.2, 6.6, 6.9
System Errors
Code Message and Corrective Action Section
900 Failed to backup setup data. Initializing will be executed.
Backup battery may be low. Maintenance service is required to replace the back-up battery.
17.7
901 Fan stopped. Power off immediately.
Maintenance service is required.
17.7
902 Backup battery is low.
Maintenance service is required to replace the back-up battery.
17.7
903 Calibration failure.
Disconnect the input and execute again.
If it fails again, service is necessary.
—
904 Invalid Command. —
17.2 Messages and Corrective Actions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL6000 Series and is the answer not in the manual?

YOKOGAWA DL6000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals