2-28
IM DLM6054-01EN
Automated Measurement of Waveform Parameters
        section 10.2 for the procedure
The DL6000/DLM6000 can perform automated measurements of various parameters on the
currently displayed acquired waveform.
Up to 100,000 results of automated measurement can be stor
ed to a file.
There are 36 different items that can be automatically measured. The DL6000/DLM6000 can
display a total of 16 measured items at a time. You can specify calculations to be made using the
results of automated measurement and display the results of those calculations.
        section 10.3 for the procedure
You can perform statistical processing on the automated measurement results described above.
You can display the five statistical processing results listed below for the values of two automated
measurement items.
Maximum (Max)
Minimum (Min)
Mean (Mean)
Standard deviation (
σ
)
The number of measured values used to calculate the statistic
s (Cnt)
Statistical processing can be performed in the following three ways.
Normal statistical processing
While acquiring signals, the DL6000/DLM6000 calculates the statistics of the specified number of
waveforms, starting with the most recently acquired waveform. If you stop waveform acquisition
and then restart it, the DL6000/DLM6000 will continue statistical processing and include the data
from before waveform acquisition was stopped. To reset statistical processing, press the Restart
soft key.
Statistical Processing within One Cycle of Measurement or the Measurement Time Range
The DL6000/DLM6000 determines periods in order from the oldest data of the displayed
waveform, measures the selected automated measurement items within each period, and
performs statistical processing on the results of automated measurement. The method used
to determine the period is the same as the method used to determine the Period waveform
parameter.
Statistical Processing of History W
aveforms
The DL6000/DLM6000 performs automated measurement on the selected range of the history
waveform and uses the results to calculate statistics. Statistical processing begins with the oldest
waveform.
2.8 Analysis and Searching

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL6000 Series and is the answer not in the manual?

YOKOGAWA DL6000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals