7-5
IM 701450-01E
7
Waveform Acquisition
Note
• When the record length is 16 MW, you cannot switch repetitive sampling mode ON/OFF. The
menu indicates OFF.
• When repetitive sampling mode is turned ON, the sample rate is set greater than or equal to 2
GS/s (greater than or equal to 5 GS/s when interleave mode is ON) for certain T/div settings.
• When repetitive sampling is turned OFF, sampling is performed at a rate of up to 1 GS/s (2
GS/s when interleave mode is ON). When the number of displayed data points is less than
500, data is interpolated and displayed. However, even when repetitive sampling is turned
OFF, repetitive sampling may be performed depending on the time axis setting.
• When repetitive sampling is in progress, waveform acquisition using the Single(N) trigger
mode is not possible.
7.4 Turning ON/OFF Repetitive Sampling Mode