5-2 IM 701450-01E
5.2 Setting V/div
<For a description of this function, refer to page 2-2.>
Procedure
SEARCH
PRESET
ZOOM
MODE
POSITION
ENHANCED
SIMPLE
LOGIC
CH 1
CH 2
CH 3
CH 4
CH 5
CH 6
CH 7
CH 8
V/DIV TIME/DIV
VERTICAL
TRIGGER
TRIG D
ACTION
DELAY
CLEAR TRACESNAPSHOT
HELP
HORIZONTAL
• To exit the menu during operation, press ESC located
above the soft keys.
• For a description of the operation using a USB keyboard or
a USB mouse, see section 4.3.
Using the V/DIV Knob
1. Press one of the CH1 to CH8/4 keys to select the channel. The CH menu
appears.
You can select up to channel CH4 and CH8 on the DL7440 and DL7480, respectively.
2. Turn the V/DIV knob to set the V/div value.
Note
• The displayed waveforms do not change if you change the V/div value while the waveform
acquisition is stopped. The new V/div value takes effect the next time the waveform
acquisition is started.
• Turning the V/DIV knob while acquisition is stopped has no affect on cursor measurement
values and automated measurement values of waveform parameters. The displayed
values are for the original V/div setting.
Using the Variable Soft Key
1. Press one of the CH1 to CH8/4 keys to select the channel. The CH menu
appears.
You can select up to channel CH4 and CH8 on the DL7440 and DL7480, respectively.
2. Press the Next (1/2) soft key
3. Press the Variable soft key.
4. Turn the jog shuttle to set the V/div value.
Note
• If you change the V/div value by turning the V/DIV knob, the Variable setting is cancelled.
• If you press the RESET key, the V/div value set using the Variable soft key is reset, and
the value returns to the initial V/div value set using the V/DIV knob.