10-72 IM 701450-01E
10.10 Performing GO/NO-GO Determination Using
Measured Waveform Parameters
Procedure
X-Y
PHASE
MENU MENU
SETUP FILE
MISC
RESET
SELECT
MATH
SHIFT
GO/NOGO
CURSOR
MEASURE
HISTORY
PRINT
ACQ
START/STOP
IMAGE SAVE
DISPLAY
• To exit the menu during operation, press ESC located
above the soft keys.
• In the procedural explanation below, the term
jog shuttle & SELECT refers to the operation of
selecting/setting items and entering values using the
jog shuttle and SELECT and RESET keys. For details
on this operation, see sections 4.1 or 4.2.
• For a description of the operation using a USB keyboard
or a USB mouse, see section 4.3.
1. Press GO/NOGO. The GO/NO-GO menu appears.
2. Press the Mode soft key. The Mode menu appears.
Entering the GO/NO-GO Determination Setup Menu
3. Press the Parameter soft key. The Parameter menu appears.
Setting the Determination Condition
4. Press the Setup soft key. The Setup dialog box opens.
5. Use jog shuttle & SELECT to set the following items of determination
parameter Param1.
• Select whether to perform determination using the corresponding
determination parameter (ON/OFF) (Mode box).
• Select a determination target waveform of CH1 to CH8/4*, Math1, or Math2
(Trace box).
• Select the measurement parameter (Item box).

Table of Contents

Other manuals for YOKOGAWA DL7440

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL7440 and is the answer not in the manual?

YOKOGAWA DL7440 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL7440
CategoryTest Equipment
LanguageEnglish

Related product manuals