2-26 IM 701450-01E
Search & Zoom
When waveform acquisition is stopped, you can search the displayed waveforms (within
the display record length. See appendix 1.) and display the waveforms that match the
search conditions expanded on the screen.
Edge Search <For the setup procedure, see section 10.4>
Search is performed by counting the number of times the waveform goes above or below
(rising or falling) a specified level.
Specified level
Displays the detected section
expanded in the zoom waveform
display frame
Search start point
Search end point
Search condition Edge: rising edge and detection count: 2
Hysteresis
Serial Pattern Search <For the setup procedure, see section 10.4>
Search is performed on whether the serial status pattern of the waveform (status pattern
of the waveform that changes over time) is the same as the status pattern set in
advance. You can also set whether the timing used to detect the waveform status (up to
64 statuses) is synchronized to a selected clock signal or is performed at certain time
intervals.
LHLL L
H
H
L
L
LL L
H
H
X
L
L
Pattern set in advance
CH2 status
Condition Clock channel: CH1, slope on which to check the status: rising,
and searched waveform: CH2
CH1
Level
CH2
Low level
High level
Detects the section that is the same as the specified status pattern
Displays the waveform expanded in the zoom waveform display frame
H: High level
L: Low level
X: Don’t care
2.6 Analyzing and Searching Waveforms