4-22 IM 701450-01E
Explanation
You can use the deskew function to minimize the effects (skew) of the delay time of the
input signals due to the probe or other factors, and observe the input signal.
You can correct the delay time of CH1 to CH8/4, Pod A, or Pod B.
You can select up to CH4 and CH8 on the DL7440 and DL7480, respectively.
Selectable Range of Skew Correction
The correction time can be set in the following range.
–100 ns to 100 ns (resolution is 0.01 ns)
4.7 Correcting the Delay Time of the Input Signals