2-38
IM DLM4038-02EN
Setting Trigger Conditions (Condition Setup)
Press the Condition Setup soft key. The screen that appears varies depending on the fast channel
data type setting.
The DLM4000 triggers on the AND of fast channel Data conditions. Items whose check boxes are
selected are used as trigger conditions.
• When the Data Type Is Nibble
Set the data pattern.
Set the data pattern input format
(Bin, Hex).
Set the comparison condition
(True, False).
• When the Data Type Is User
Select the check boxes for the items that you want to use as comparison conditions.
Set the comparison condition (Data = a; Data ≠ a; a ≤ Data;
Data ≤ b; a ≤ Data ≤ b; Data<a, b<Data.
Set reference values a
and b.
Every Slow CH Mode
Press the Mode soft key and then the Every Slow CH soft key to display the following menu.
The DLM4000 triggers when it detects an “Every Slow CH” message.
Set the slow channel message type (Short, Enhanced).
This is selectable when the format version is JAN2010.
2.13 Triggering on SENT Signals (Option)