3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 5.2 Tube Detect Sensor (Beam Block) Adjustment
3 - 72
3
Search
Book TOC
Go Back
Figure 3-33 Black Beam Block Mechanical Adjustment
5. Perform Electrical Adjustment as follows:
• [NON-CE MARK]
Perform
Steps 6
and
7
.
• [CE MARK]
Perform
Step 8
.
Loosen
Set Screw
(1.5mm Allen)
SL-187
Rotate
Beam
Block