VP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment
3 - 119
3
Search
Book TOC
Go Back
f. Connect Mixin
Motor Current Tool to Mixin
Motor #2 (Fi
ure 3-63, Pa
e-3-120).
. Press
R
. Loader should beep.
h. If no beep, turn R - 84 Clockwise until Loader beeps.
i. Reconnect Mix Motor # 2.
j. Go to
Step 11
(Pa
e 3-115). Procedure is complete.