3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 13.1 Lead Bubble Check/Adjustment
3 - 157
3
Search
Book TOC
Go Back
9. If Incorrect:
E Loader with old (clear plastic) Wash Block.
Install new (brown plastic) Wash Block PL-5.13 or,
Adjust Wash block until lead bubble is correct.
- Place shim (
.020" (0.5mm)) under left side of wash block near mountin
g
screw.