CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 13.1 Lead Bubble Check/Adjustment
3 - 158
3
Search
Book TOC
Go Back
- Adjust shim to achieve conditions at top and bottom of Wash block, Fi
g
ure 3-84
E Loader with new (brown plastic) Wash Block.
Replace Wash Block.
F, G Loader.
No adjustment..
Figure 3-84 E Loader Sample Needle Alignment
--- END ---
SL-232
Needle
Touches
Here
Needle
Almost
Touches
Here
Needle Hole
Centered
on Waste
Drain Cavit
y