28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 15.1 Barcode Reader Configuration (Europe Only)
3 - 166
3
Search
Book TOC
Go Back
- Press
[ENTER]
.
Figure 3-87 Barcode Default Settings
13.To save new settin
g
s:
CURRENT SETTINGS FOR CODE TYPES
------------------------------------------------------------------------
| CODE 39 | CODABAR | I 2 OF 5 | UPC | CODE 128 |
----------------|----------|----------|----------|----------|----------|
CODE TYPE | ENABLED | ENABLED | ENABLED | DISABLED | DISABLED |
FIXED LENGTH | DISABLED | DISABLED | N/A | N/A | DISABLED |
CODE LENGTH #1 | 10 | 10 | 12 | N/A | 10 |
CODE LENGTH #2 | N/A | N/A | 10 | N/A | N/A |
CHECK DIGIT | DISABLED | DISABLED | DISABLED | N/A | N/A |
C/D OUTPUT | DISABLED | DISABLED | DISABLED | N/A | N/A |
INTERCHAR GAP | DISABLED | DISABLED | N/A | N/A | N/A |
S/S MATCH | N/A | DISABLED | N/A | N/A | N/A |
S/S OUTPUT | N/A | DISABLED | N/A | N/A | N/A |
EAN | N/A | N/A | N/A | DISABLED | N/A |
SUPPLEMENTALS | N/A | N/A | N/A | DISABLED | N/A |
NARROW MARGINS | DISABLED | DISABLED | DISABLED | DISABLED | DISABLED |
------------------------------------------------------------------------
ESC = MAIN MENU OR EXIT N = NEXT ITEM
M = PREVIOUS MENU SP = NEXT ITEM
B = PREVIOUS ITEM CR = THIS ITEM