6241A/6242 DC Voltage Current Source/Monitor Operation Manual
5.2.3 Sweep Source Mode Operation
5-16
The output value before sweeping starts is a bias value.
When start trigger is input, the start value is output.
When the Hold time passes after the start trigger, sweeping starts.
If trigger mode is AUTO, sweep step changes after the period time. However, if measurement has
not completed, the next step is delayed until the measurement is completed.
If the trigger mode is HOLD, sweep step is enabled whenever the trigger is input.
Measure-
ment range
changed
AUTO
Measurement
range changes
while sweeping
Th: Hold Time
Tp: Period Time
Td: Measurement
Delay Time
Tds:Source Delay Time
Tm: Measurement Time
(Integration time +
Measurement data
processing time)
Tcn: Operation Processing Time
Trc: Range Change Processing
Time
HOLD
Table 5-5 DC Sweep Source Mode (2/2)
Operational
condition
Trigger mode Explanation Operation Remarks
Tp
Auto range processing
Th
Tp
Tp
Td
Tm
Td
Tm
Tds
Tds
Tds
1 1’ 1”
2
3 3’
Start trigger
Bias
Operate
TRIGGER IN
COMPLETE OUT [HI/GO/LO]
SYNC OUT
COMPLETE OUT [END]
COMPLETE OUT [FRONT]
1’
Bias
Th
Td
Tm
Td
Tm
Tds
1
2
Start trigger
Operate
Trigger
TRIGGER IN
SYNC OUT
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
COMPLETE OUT [FRONT]
Auto range
processing
Tp
Tds
Tp