6241A/6242 DC Voltage Current Source/Monitor Operation Manual
A-2
3. Measurement value is
not being output.
Cause:
Solution:
Unit in Standby or Suspended status.
Set to Operate and verify that the OPR indicator is ON.
Cause:
Solution:
Measurement is not ON.
Verify measurement ON/OFF setting.
Cause:
Solution:
When measuring in auto range, the value is unstable and the range
unconfirmed, therefore measurement data is not output.
Change to a fixed range and measure.
Cause:
Solution:
A trigger signal is not input even when the trigger signal cable is con-
nected to the external trigger.
Verify TRIG INPUT connection cable and the signal.
4. A source value or
measurement value
indication is unstable
or is in error.
Cause:
Solution:
Function or range settings have an error.
Verify the setting again.
Cause:
Solution:
Cable connection is wrong.
Verify cable connections again.
Cause:
Solution:
The cable is disconnected.
Verify the cables with the tester. If in error, replace it.
Cause:
Solution:
The cable is connected to a wrong terminal.
Verify cable connections again.
Cause:
Solution:
The induction noise scatters the measurement value.
Set the integration time to 1 PLC or over.
5. The measurement
value is over range.
Cause:
Solution:
When the NULL calculation value becomes twice or more of the value
of full-scale.
Raise the source value or limiter range.
6. Unable to input set-
tings with the mea-
surement control key.
Cause:
Solution:
While inputting with direct mode, the set value is at half-brightness and
only the green keys on the panel are enabled.
Press 123... key to complete the direct input mode.
7. A relay life error
occurred.
Cause:
Solution:
The Operate relay operated more than one million time.
In this case, contact an ADC CORPORATION sales representative.
Table A-1 Items to be Inspected before Requesting the Repair (2/2)
Q (Symptom) A (Cause and Solution)