ABORt
ABORt[:ALL]
Function Stops all measurements that are active.
If the trigger arm is set to single, the measurements will go to the idle state.
If the trigger arm is set to continuous, the measurements will re-arm and initiate again.
Setting Range
AFANalyzer - AF Analyzer measurement
ATXPower - Analog Transmitter Power measurement
CAPPower - Access Probe Power measurement
CCTPhase - Code Channel Time/Phase error measurement
CPOWer - Channel Power measurement
CFERror - Frame Error Rate measurement
CTXSpurious - TX Spruious Emmissions measurement
FSTability - Frequency Stability measurement
FM - Frequency Modulation measurement
GPOWer - Gated Power measurement
HWQuality - Handoff Waveform Quality measurement
SAUDio - Swept Audio measurement
TFERror - TDSO Frame Error Rate measurement
WQUality - Waveform Quality measurement
Programming Example
OUTPUT 714;"ABORT:ALL" !Aborts all active measurements in progress.