S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_aparameter.fm
CALL[:CELL]:APARameter
CALL[:CELL]:APARameter:POWer:NOMinal[:OFFset]
CALL[:CELL]:APARameter:POWer:NOMinal:EXTended
Function Sets/queries the nominal power offset.
This offset value serves as a correction factor to the open loop power control equation.
Setting Range: -8 to 7 dBm with nominal power extended set to 0 (false). See
CALL[:CELL]:APARameter:POWer:NOMinal:EXTended .
Resolution: integer
Query Range: -8 to 7 dBm with nominal power extended set to 0 (false). See
CALL[:CELL]:APARameter:POWer:NOMinal:EXTended .
Resolution: integer
*RST Setting 0
Programming Example
OUTPUT 714;"CALL:CELL:APARAMETER:POWER:NOMINAL:OFFSET -5"
!Sets the nomiinal power offset to -5 dB.
Function Sets/queries extended nominal power.
Setting this parameter to 1 (true) introduces a -16 dB offset to the open loop power control
correction range. The extended nominal power range is from -24 dB to -9 dB inclusive.
With this parameter set to 0 (false) the open loop power control correction range is -8 dB to 7 dB.
See CALL[:CELL]:APARameter:POWer:NOMinal[:OFFset] .
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0
Programming Example
OUTPUT 714;"CALL:CELL:APARAMETER:POWER:NOMINAL:EXTENDED ON"