360
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_qpchannel.fm
CALL:QPCHannel
CALL:QPCHannel:STATe
Function Sets/queries the forward quick paging channel’s (QPCH) state. When the state of the forward
QPCH channel is on, the test set will transmit it even if the mobile station does not support
QPCH.
The QPCH channel is slotted (pulsed). The test set adjusts OCNS (orthogonal channel noise
simulator) to an appropriate level for the periods of time when the slotted QPCH signal is pulsed
on. When QPCH is pulsed off, the output level will not be fully calibrated.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"CALL:QPCHANNEL:STATE ON" !Turns on the quick paging channel.