CAL L: S TATus
CALL:STATus:CELL:POWer:STATe[:SELected]?
CALL:STATus:CELL:SYSTem[:TYPE]?
CALL:STATus:FCHannel:LEVel[:SELected]?
Function Queries the state of the cell power measurement for the currently selected system type.
Cell power state is set using the CALL[:CELL]:POWer commands.
The optional [:SELected] keyword in this command specifies that the power state being queried
applies to the current system type.
Query Range: 1|0
*RST Setting 0 (Off)
Programming Example
OUTPUT 714;"CALL:STATUS:CELL:POWER:STATE:SELECTED?"
Function Queries the current system type that the test set is currently emulating.
When the test set is in active cell operating mode (see CALL[:CELL]:OPERating ) the current
system type is selected by the CALL[:CELL]:SYSTem command.
When the test set is in any test mode operating mode (see CALL[:CELL]:OPERating ) or CW
operating mode this command will return the corresponding system type.
Query Range: DIG2000|CW|AMPS|DIG95
*RST Setting DIG2000
Programming Example
OUTPUT 714;"CALL:STATUS:CELL:SYSTEM:TYPE?"
Function Queries the current forward fundamental channel (FCH) level.
The forward fundamental channel level is set using the CALL:FCHannel commands.
The optional [:SELected] keyword in this command specifies that the fundamental channel level
being queried applies to the current system type.
Query Range: 0 to 30 dB, 9.91 E+37
Resolution: 0.01 dB
*RST Setting 9.91 E+37
Programming Example
OUTPUT 714;"CALL:STATUS:FCHANNEL:LEVEL?"