397
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_status.fm
CAL L: S TATus
CALL:STATus:SCHannel:STATe[:SELected]?
CALL:STATus:SCHannel:SYNChronized?
CALL:STATus:SYNC:LEVel[:SELected]?
CALL:STATus:SYNC:STATe[:SELected]?
Function Queries the current on/off state of the forward supplemental channel (SCH).
The forward supplemental channel on/off state is set using the CALL:SCHannel commands.
The optional [:SELected] keyword in this command specifies that the supplemental channel state
being queried applies to the current system type.
Query Range: 1|0
*RST Setting 0 (Off)
Programming Example
OUTPUT 714;"CALL:STATUS:SCHANNEL:STATE?"
Function Queries the if the data on the F-SCH is synchronized when in SO32 (TDSO). This is primarily
used by the TDSO FER measurement. The returned setting is NSCH (no F-SCH assignedl),
NSYN (not synchronized), or SYNC (synchronized.
The forward supplemental channel on/off state is set using the CALL:SCHannel commands.
Query Range: NSCH|NSYN|SYNC
*RST Setting NSCH
Programming Example
OUTPUT 714;"CALL:STATUS:SCHANNEL:SYNCHRONIZED"
Function Queries the current forward sync channel level for the currently selected system type.
The forward sync channel level is set using the CALL:SYNC commands.
The optional [:SELected] keyword in this command specifies that the sync level being queried
applies to the current system type.
Query Range: 0 to -20 dB, 9.91 E+37
Resolution: 0.01 dB
*RST Setting 9.91 E+37
Programming Example
OUTPUT 714;"CALL:STATUS:SYNC:LEVEL:SELECTED?"
Function Queries the current sync channel level state for the currently selected system type.
The forward sync channel level on/off state is set using the CALL:SYNC commands.
The optional [:SELected] keyword in this command specifies that the sync state being queried
applies to the current system type.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals