405
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_total_power.fm
CALL:TOTal:POWer
CALL:TOTal:POWer[:AMPLitude][:SELected]?
CALL:TOTal:POWer:STATe[:SELected]?
Function Queries the total RF power settings for the current system type.
Total RF power is the sum of cell power level and AWGN power level.
The return value for this query is not necessarily the power level being transmitted by the test
set at the moment this command is processed by the test set. It may return a value that the test
set has been requested to transmit, but that has been pended or delayed for some reason. To
query the power level the test set is transmitting at the moment the command is sent, use the
“CALL:STATus:TOTal:POWer[:AMPLitude][:SELected]?” command.
The optional [:SELected] keyword in this command specifies that this query applies to the
current system type.
Query Range: -170 dBm/1.23 MHz to +35 dBm/1.23 MHz, 9.91 E+37
Resolution: 0.01 dB
*RST Setting 9.91 E+37
Programming Example
OUTPUT 714;"CALL:TOTAL:POWER:AMPLITUDE:SELECTED?"
Function Queries the total RF power state for the current system type.
The total RF power state is 1 (on) if either cell power or AWGN power is on.
The return value for this query is not necessarily the total power state at the moment this
command is processed by the test set. This command may return a state that has been requested
of the test set, but that has been pended or delayed for some reason.To query the power state of
the test set at the moment, use the “CALL:STATus:TOTal:POWer:STATe[:SELected]?” command.
The optional [:SELected] keyword in this command specifies that this query applies to the
current system type.
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"CALL:TOTAL:POWER:STATE:SELECTED?"