S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_fstability.fm
FETCh:FSTability
FETCh:FSTability[:ALL]?
FETCh:FSTability:FERRor[:WORSt]?
Function Queries the full results of a frequency stability measurement. Returns integrity, worst-case frequency
error, and average frequency.
Query Worst-case frequency error:
Range: -500 to 500 parts per million (ppm)
Resolution: 0.01 ppm
Average frequency:
Range: 100 E+6 to 3 E+9 Hz
Resolution: 0.01 ppm
Programming Example
OUTPUT 714;FETCH:FSTABILITY:ALL?
!Returns integrity, worst-case frequency error, and average frequency.
Function Queries the worst case frequency stability measurement during a multi-measurement.
Query Range: -500 to 500 parts per million (ppm)
Resolution: 0.01 ppm
Programming Example
OUTPUT 714;FETCH:FSTABILITY:FERROR:WORST?