449
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_fstability.fm
FETCh:FSTability
FETCh:FSTability:FERRor:ALL?
FETCh:FSTability:FERRor:AVERage?
Function Queries the full results of a frequency stability .
Measurement results are returned in the following order: minimum frequency error, maximum
frequency error, average frequency error, worst-case frequency error.
Query Maximum, minimum frequency error:
Range: -500 E+3 to 500 E+3 Hz
Resolution: 1 Hz
Average frequency error:
Range: -500 E+3 to 500 E+3 Hz
Resolution: 1 Hz
Worst-case frequency error:
Range: -500 to 500 parts per million (ppm)
Resolution: 0.01
Programming Example
OUTPUT 714;FETCH:FSTABILITY:FERROR:ALL?
!Returns minimum, maximum, average, and worst-case frequency error.
Function Queries the average frequency error measurement.
Query Range: -500 E+3 to 500 E+3 Hz
Resolution: 1 Hz
Programming Example
OUTPUT 714;FETCH:FSTABILITY:FERROR:AVERAGE?

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals