479
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_initiate.fm
INITiate
INITiate:ON?
INITiate:<measurement mnemonic>:OFF
INITiate:COUNt?
INITiate:DONE?
Function Queries the measurements that are on. Returns a comma separated list of measurement
mnemonics.
Range AFAN|ATXP|CAPP|CCTP|CFER|CPOW|CTSX|DAP|FM|FST|GPOW|HWQ|SAUD|
TFER|WQU|NONE
Programming Example
OUTPUT 714;"INITIATE:ON?"
Function Deactivates the selected measurement. See “Measurement States” on page 220.
Only one measurement can be deactivated at a time. To stop one or more measurements and
leave them in the active state, use “ABORt” on page 251.
Programming Example
OUTPUT 714;"INITIATE:CPOWER:OFF" !Deactivates the channel power measurement.
Function Queries the number of measurements that have been initiated (are active). See “Measurement
States” on page 220.
Query Range: 0 to 10
Programming Example
OUTPUT 714;"INITIATE:COUNT?"
Function Queries (one at a time) which measurements, if any, are finished or have timed-out.
See “Measurement Event Synchronization” on page 231 for how to use this command.
See “Measurement States” on page 220 to understand the test set’s measurement states.
See “INITiate:DONE:FLAG:<measurement mnemonic>” on page 480 for include or exclude
commands.
Range AFAN|ATXP|CAPP|CCTP|CFER|CPOW|CTSX|DAP|FM|FST|GPOW|HWQ|SAUD|
TFER|WQU|NONE|WAIT