522
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cctphase.fm
SETup:CCTPhase
SETup:CCTPhase:COUNt:STATe
SETup:CCTPhase:PERRor[:LIMit]
SETup:CCTPhase:TERRor[:LIMit]
Function Sets/queries the code channel timing and phase error multi-measurement state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:CCTPHASE:COUNT:STATE ON" !Turns on code channel timing and phase error
multi-measurements.
Function Sets/queries the code channel phase error limit.
Setting Range: 0 to 1.00 radian
Resolution: 0.01 radian
Query Range: 0 to 1.00 radian
*RST Setting 0.15 radian
Programming Example
OUTPUT 714;"SETUP:CCTPHASE:PERROR 0.25"
Function Sets/queries the code channel timing error limit.
Setting Range: 0 to 100.0 ns
Resolution: 0.1 ns
Query Range: 0 to 100.0 ns
*RST Setting 10.0 ns
Programming Example
OUTPUT 714;"SETUP:CCTPHASE:TERROR 25"