SETup:CFERror
SETup:CFERror:TIMeout[:STIMe]
SETup:CFERror:TIMeout:STATe
SETup:CFERror:TIMeout:TIME
Function The setting form of this command performs two functions: (1) sets the FER (frame error rate)
measurements timeout value, and (2) sets the FER measurements timeout STATe to 1 (on).
The query form returns the FER measurements timeout value.
Setting Range: 1 to 200,000 seconds
Resolution: 1 second
Query Range: 1 to 200,000
*RST Setting 200 seconds, on
Programming Example
OUTPUT 714;"SETUP:CFERROR:TIMEOUT:STIME 200S
!Sets timeout state to ON and sets the timeout time to 200 seconds.
Function Sets/queries the FER (frame error rate) measurements timeout state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:CFERROR:TIMEOUT:STATE ON
Function Sets/queries the FER (frame error rate) measurements timeout setting to be used when the
timeout state is set to ON.
Setting Range: 1 to 200,000 seconds
Resolution: 1 second
Query Range: 1 to 200,000
*RST Setting 200 seconds
Programming Example
OUTPUT 714;"SETUP:CFERROR:TIMEOUT:TIME 200S

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals