530
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cpower.fm
SETup:CPOWer
SETup:CPOWer:COUNt:STATe
SETup:CPOWer:MSPeed
Function Sets/queries the channel power multi-measurement state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:CPOWER:COUNT:STATE ON" !Turns on channel power multi-measurements.
Function Sets/queries the channel power measurement speed.
Channel power measurement speed is determined by the measurement sampling period.
NORMal channel power measurements are sampled for 10 milliseconds, while FAST channel
power measurements are sampled for 1.25 milliseconds.
NORMal speed channel power measurements are compatible with both gated (RF) or non-gated
signals. Since NORMal channel power measurements are slower than FAST measurements, use
NORMal when measuring gated input signals. Gated input signals occur when radio
configuration 1 or 2 are selected and any data rate other than full is selected. See
“CALL:TRAFfic” and “CALL[:CELL]:RCONfig” .
FAST channel power measurement speed is compatible only with non-gated (RF) signals.
Non-gated signals occur when radio configuration 3, 4, or 5 are selected or radio configuration 1
or 2 is selected with full rate. See “CALL:TRAFfic” and “CALL[:CELL]:RCONfig” .
Setting Range: NORMal|FAST
Query Range: NORM|FAST
*RST Setting NORM
Programming Example
OUTPUT 714;"SETUP:CPOWER:MSPEED FAST" !Sets up fast channel power measurements.