S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_tferror.fm
TDSO FER (Frame Error Rate) Troubleshooting
TDSO FER (Frame Error Rate) Troubleshooting
TDSO FER Setup Issues
TDSO FER results may take up to 20.48 seconds to display
A call must be setup using SO32 (+F-SCH) and that call must be currently connected.
If No F-SCH is displayed in the bottom right portion of the display, the supplemental channel is not
connected. If No Sync is displayed, the sync channel has not been sent to the mobile station yet.
The TDSO FER test returns the confidence level based on the actual average FER result obtained over the
fixed frame count length. If the confidence level is less than 95% for the specified FER confidence requirement,
the test has not passed minimum standards.
Integrity indicators do not apply to this measurement.
Related Topics
TDSO Frame Error Rate Measurement Description

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals