726
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_err_msg_neg400.fm
-400 through -499 Error Message Descriptions
-400 through -499 Error Message Descriptions
Description
A Query error is generated either when data in the instrument’s GPIB output queue has been lost, or when an
attempt is being made to read data from the output queue when no output is present or pending.
Error Message Description
-400 Query error This event bit (Bit 2) indicates that an attempt to read
data from the Output Queues when no output is
present or pending, to data in the Output Queue has
been lost see IEEE488.2, 11.5.1.1.7.
-410 Query INTERRUPTED Indicates the test set has been interrupted by a new
program message before it finishes sending a
RESPONSE MESSAGE see IEEE 488.2, 6.3.2.3.
-420 Query UNTERMINATED Indicates an incomplete Query in the program see
IEEE 488.2, 6.3.2.2.
-430 Query DEADLOCKED Indicates that the Input Buffer and Output Queue are
full see IEEE 488.2, 6.3.1.7.
-440 Query UNTERMINATED after indefinite response Indicates that a query was received in the same
program message after a query requesting an
indefinite response was executed see IEEE 488.2,
6.5.7.5.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals