EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #771 background imageLoading...
Page #771 background image
771
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_conf_timebase.fm
Timebase Description/Configuration
Timebase Description/Configuration
Description
The time base source is selected by the test set, either an internal time base or an external source (if a suitable
signal is detected) is used as the reference oscillator. If a 10 MHz +/- 100 ppm signal, that has an input level
from 0 to +13 dBm is connected to the 10 MHz REF IN connector on the rear panel, the test set will
automatically select the external timebase.
You can read the status window at the bottom of the test set display for the EXT REF indicator, or query the
test set to verify if it is using an external time base or an internal time base. You may also query the test set to
verify if the time base is locked. The reference oscillator functionality is controlled through the SYSTEM
subsystem.
Example:
OUTPUT 714;SYSTEM:ROSCILLATOR[:TIMEBASE]? !returns INT or EXT
!(internal or external) timebase.
OUTPUT 714;SYSTEM:ROSCILLATOR:LOCKED? !returns 1 or 0 (locked or unlocked)
!condition for timebase
Related Topics
SYSTem:ROSCillator
Rear Panel Connectors

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals