EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
769 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 125
Calibration
Full 4-Port Calibration
4. Calibration
Figure 4-20 Connecting standards in full 4-port calibration
Procedure
Step 1. Press or to select the channel for which you want to perform the
calibration.
Step 2. Press .
Step 3. Press Calibrate.
Step 4. Press 4-Port Cal.
Step 5. Press Reflection.
Step 6. Connect an OPEN calibration standard to test port 1 (the connector to which the DUT is to
be connected).
Step 7. Press Port 1 Open to start the calibration measurement.
Step 8. Disconnect the OPEN calibration standard connected in Step 6 and replace it with a
SHORT calibration standard.
Step 9. Press Port 1 Short to start the calibration measurement.
Step 10. Disconnect the SHORT calibration standard connected in Step 8 and replace it with a
LOAD calibration standard.
Step 11. Press Port 1 Load to start the calibration measurement.
Step 12. Repeat Step 6 to Step 11 on test port 2.

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals