EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
769 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 235
Calibration
Scalar-Mixer Calibration
4. Calibration
Figure 4-75 Status indication for the traces in which scalar-mixer is valid
NOTE In scalar-mixer calibration, the normal calibration coefficient is invalid while the
frequency-offset sweep is in progress; in this case the scalar-mixer calibration coefficient is
used instead. Turning the frequency-offset sweep (
Frequency Offset) OFF switches over
to the normal calibration coefficient; however, the information on the scalar-mixer
calibration coefficient is retained.
Operational Procedure (when using mechanical calibration kit)
NOTE Before starting scalar-mixer calibration, verify that the frequency-offset (Frequency
Offset) is ON. If the frequency-offset is not valid, you cannot start scalar-mixer calibration.
Step 1. Press or to select the channel you want to calibrate.
Step 2. Press .
Step 3. Press Mixer/Converter Calibration.
Step 4. Press Scalar Cal (Manual).
Step 5. Press Select Ports.
Step 6. Select the test port from the list below for starting scalar-mixer calibration. In the following
steps, x and y signify the selected test ports.
Softkey Function
Port 1 Selects test port 1
Port 2 Selects test port 2
Port 3 Selects test port 3
Port 4 Selects test port 4
2-1 (fwd) Selects forward direction for test ports 1 and 2
1-2 (rev) Selects reverse direction for test ports 1 and 2
1, 2 (both) Selects both directions for test ports 1 and 2
3-1 (fwd) Selects forward direction for test ports 1 and 3
1-3 (rev) Selects reverse direction for test ports 1 and 3
1, 3 (both) Selects both directions for test ports 1 and 3
4-1 (fwd) Selects forward direction for test ports 1 and 4

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals