204 Chapter 4
Calibration
Specifying Different Standard for Each Frequency
Disabling standard defined for a subclass
The following procedure shows how to disable a standard defined for a subclass. Note that
you cannot disable subclass 1 because at least one standard must exist. The following
procedure shows how to disable an OPEN standard of subclass 2.
Procedure
Step 1. Press .
Step 2. Press Cal Kit.
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of “Defining different standard for each frequency band” on page 202.)
Step 4. Press Modify Kit.
Step 5. Press Specify CLSs.
Step 6. Press Sub Class, and select Sub Class 2.
Step 7. Select the standard type Open, press Set All, make settings for all ports.
Step 8. Select None.
Step 9. Press Return.