Chapter 12 403
Optimizing Measurements
Setting Port Extensions and Loss Values
12. Optimizing Measurements
2. Setting a DC loss value
Step 1. Press or to activate the channel for which you want to set a DC
loss value.
Step 2. Press .
Step 3. Press Port Extensions.
Step 4. Press Loss.
Step 5. Press Select Port to select the port for which you want to set a DC loss value.
Step 6. Press Loss at DC, and enter a DC loss value.
Step 7. If you want to set a DC loss value for other ports, repeat Step 5 to Step 6.
Enabling port extensions and loss values
Step 1. Press or to activate the channel for which you want to enable
port extensions and loss values.
Step 2. Press .
Step 3. Press Port Extensions.
Step 4. Turn on Extensions.