Chapter 11 381
Limit Test
Limit Test
11. Limit Test
Limit line offset
By adding a certain offset to the limit value, you can adjust the limit line so that it conforms
to the device output.
Step 1. Press or to activate the channel to which you want to apply the
limit test function.
Step 2. Press or to activate the trace to which you want to apply the limit test
function.
Step 3. Press to display the Analysis menu.
Step 4. Press Limit Test to display the softkeys for the limit test.
Step 5. Press Limit Line Offsets to display the limit line offset function menu. The following
functions correspond to each softkey.
Softkey Function
Stimulus Offset Adds a certain offset to the stimulus value of the entire segment in the
limit table. (Stimulus offset)
Amplitude Offset Adds a certain offset to the response value of the entire segment in the
limit table. (Amplitude offset)
Marker Amp. Ofs. Adds the amplitude offset by the same amount as the retrieved value of the
active marker. You can confirm the current value set for the amplitude
offset by pressing
Amplitude Offset. (Marker amplitude offset)
Return Ends the limit line offset menu.
Figure 11-5 Stimulus offset