Chapter 4 203
Calibration
Specifying Different Standard for Each Frequency
4. Calibration
Defining standard for each subclass
This section demonstrates the procedure to specify a different OPEN standard for each
frequency band, using subclasses #1 and #2. In this example, the standards created in
“Defining different standard for each frequency band” on page 202 are used.
Procedure
Step 1. Press .
Step 2. Press Cal Kit.
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of “Defining different standard for each frequency band” on page 202.)
Step 4. Press Modify Kit.
Step 5. Press Specify CLSs.
Step 6. Press Sub Class, and select Sub Class 1.
Step 7. Select the standard type Open, press Set All, make settings for all ports.
Step 8. Select #1 (Open 3G).
Step 9. Press Return.
Step 10. Then, press Sub Class, and select Sub Class 2.
Step 11. Select the standard type Open, press Set All, make settings for all ports.
Step 12. Select #2 (Open 6G).
Step 13. Press Return.